F002enter a password“+ - - +(Turn the knob)”,Then press the knob to enter; F001->F100enter a password“+ - + -(Turn the knob)”,Then press the knob to enter; Press "Ok" to enter the setting of corresponding parameter value, and turn the knob to modify parameters; Press "Ok" to save the parameter, exit the parameter setting, and display the next parameter number (after setting the F001 value without entering the password, the normal operation screen will enter); Rotate the knob directly to the last parameter number, press the "OK" key, and the setting is completed; If no operation on the parameter setting screen exceeds 10s, it will exit the setting state; Voltage setting parameters. please read section 3.4 before operating, especially to enter a password: Number Description Note Default F001 Adjust Mode 0:Coarse 1:Fine 2: Point Mode(Need to set F100>0) 0 F002 Output Mode 0:±10V 1:±5V 2:0-10V 3:2-10V 4:0-5V 5:1-5V 6:0-3.3V 7:0-2.5V 8:0-1V 9:-10V-0V 2 F003 Display Mode 0:Real Voltage 1:Percentage 0-100.0% 2:50HZ 3:1500 0 F004 Add Or Sub Num For Knob’s Pulse(Coarse) 1-50 No Decimal Point Concept (1-50)×10 1 F005 Add Or Sub Num For Knob’s Pulse ( Fine) 1-50 No Decimal Point Concept (1-50)×1 1 F006 -10V Calibration Value -199 -- +199 Internal Reference,Please Be Careful F007 0V Calibration Value -199 -- +199 Internal Reference,Please Be Careful F008 +10V Calibration Value -199 -- +199 Internal Reference,Please Be Careful F100 Point Mode Num 0: Point Mode Not Use 2-9: Point Num 0 F101…F109 Point Output Value range : -10V to 10V You can set as many values as there are points Current setting parameters. please read section 3.4 before operating, especially to enter a password: Number Description Note Default F001 Adjust Mode 0:Coarse 1:Fine 2: Point Mode(Need to set F100>0) 0 F002 Output Mode 0:0-20mA 1:4-20mA 2:0-22mA 0 F003 Display Mode 0:Real Voltage 1:Percentage 0-100.0% 2:50HZ 0 F004 Add Or Sub Num For Knob’s Pulse(Coarse) 1-50 No Decimal Point Concept (1-50)×10 1 F005 Add Or Sub Num For Knob’s Pulse ( Fine) 1-50 No Decimal Point Concept (1-50)×1 1 F008 Calibration Value -999 -- +999 Internal Reference,Please Be Careful F100 Point Mode Num 0: Point Mode Not Use 2-9: Point Num 0 F101…F109 Point Output Value range : 0-22mA You can set as many values as there are points current problem detection: the possibility of the current not output is very small,you can measured the voltage of 4-AIO 6-GND with a multimeter, if it is >20V, than current file of your multimeter is broken or gear is wrong, because there are voltage must be current, try change a multimeter, multimeter current file inside a fuse is easy to burn; Please Read This Manual Carefully Before Wiring,Turn Off The Power And Then Connect; ">Aideepen High-precision handheld signal generator without battery voltage ±0-5-10V

Aideepen High-precision handheld signal generator without battery voltage ±0-5-10V

MODEL:QH-VISG2-ED(With battery) / QH-VISG2-EN(No battery)

Voltage And Current Signal Generator-User manual V202008

Technical Indicators


	three power supply modes ,using the mobile phone charger or external 24V can charge/work;
	
		External wiring terminal power supply DC15-30V 1W(without charging)/4W(charging);
		External MicroUSB-5V current 0.2a (no charge)/0.8A(charge);
		(Optional) The built-in 3.7V lithium battery(1000mah), the maximum charging current is 0.6A,Generally, it takes more than 4 hours to fully charge. It can be used for 1-5 hours;
	
	
	The voltage output is adjustable ± 10V, Error
	The current output is adjustable 0-22ma,Error
	working environment: 0-40 ℃, relative humidity 
	Size (cm):Shell=100x60x20 ,The plug protruding 9cm,The knob protruding 15cm


Wiring diagram






	
		
		1
		
		
		GND
		
		
		Power supply negative (input)
		
	
	
		
		2
		
		
		24V
		
		
		Power supply positive (input)
		
	
	
		
		3
		
		
		AI+
		
		
		Current passive mode 24V+ input
		
	
	
		
		4
		
		
		AIO
		
		
		Current signal output
		
	
	
		
		5
		
		
		AVO
		
		
		Voltage signal output
		
	
	
		
		6
		
		
		GND
		
		
		Signal ground (connect 1-GND)
		
	




Voltage signal output:5-AVO  6-GND
Common active mode of current output:Our output 4-AIO6-GND corresponds to PLC I + I -!
Current passive mode:The PIN 3-AI+ is externally connected to 24V+, and then output from 4-AIO to control the external 24V current.Our meter is equivalent to an adjustable resistance;Therefore, it is called passive mode.



Usage:


	Operation panel and operation instructions







Switching - Adjust the display V or I, press the "MODE" key to switch and adjust the display V or I MODE

The indicator light shows what state it is in and what mode the system operation is in;

The original output will continue to output After switching, but not in the adjustment status;

Turn on the battery switch to check the battery status and quantity

Batterystatus indicator:




	
		
		Battery indicator
		
		
		Green
		
		
		Yellow
		
		
		Red
		
		
		Twinkle
		
		
		Green not Twinkle (External power)
		
	
	
		
		capacity
		
		
		≥80%
		
		
		≥40%
		
		
		>0
		
		
		Charging。。。
		
		
		Full
		
	




Hold the "MODE" key for 1 second and the digital tube flashes for 5 seconds to display the battery voltage;



Digital adjustment and parameter setting:

(Forewarned:Push Down The Knob Is “OK”,Clockwise Rotation The Knob One Pulse Is” + ”, Anticlockwise Rotation The Knob One Pulse Is” - ”)

Short Press “OK” To Save The Output Value,Digital Tube Display "..." After loosening, Saved Successfully, The Output Value Is This Saved Value After The Next Boot; When We Debug The Device, We Need To Adjust The Output Value Randomly, As Long As The Knob Is Not Pressed, It Will Still Be The Saved Value After Rebooting;
Hold The “OK” For 2 Seconds To Enter The Setting State.,Display”F001”(Parameter Number 001);
Turn the knob to change the parameter number, You need to enter a password when enter F002 or F100:

	F001->F002enter a password“+ - - +(Turn the knob)”,Then press the knob to enter;
	F001->F100enter a password“+ - + -(Turn the knob)”,Then press the knob to enter;


Press "Ok" to enter the setting of corresponding parameter value, and turn the knob to modify parameters;
Press "Ok" to save the parameter, exit the parameter setting, and display the next parameter number (after setting the F001 value without entering the password, the normal operation screen will enter);
Rotate the knob directly to the last parameter number, press the "OK" key, and the setting is completed;
If no operation on the parameter setting screen exceeds 10s, it will exit the setting state;



Voltage setting parameters. please read section 3.4 before operating, especially to enter a password:




	
		
		Number
		
		
		Description
		
		
		Note
		
		
		Default
		
	
	
		
		F001
		
		
		Adjust Mode
		
		
		0:Coarse  1:Fine  2: Point Mode(Need to set F100>0)
		
		
		0
		
	
	
		
		F002
		
		
		Output Mode
		
		
		0:±10V  1:±5V  2:0-10V  3:2-10V  4:0-5V  5:1-5V  6:0-3.3V  7:0-2.5V  8:0-1V  9:-10V-0V
		
		
		2
		
	
	
		
		F003
		
		
		Display Mode
		
		
		0:Real Voltage  1:Percentage 0-100.0%  2:50HZ  3:1500
		
		
		0
		
	
	
		
		F004
		
		
		Add Or Sub Num For Knob’s Pulse(Coarse)
		
		
		1-50

		No Decimal Point Concept  (1-50)×10
		
		
		1
		
	
	
		
		F005
		
		
		Add Or Sub Num For Knob’s Pulse ( Fine)
		
		
		1-50

		No Decimal Point Concept  (1-50)×1
		
		
		1
		
	
	
		
		F006
		
		
		-10V Calibration Value
		
		
		-199 -- +199  Internal Reference,Please Be Careful
		
		
		 
		
	
	
		
		F007
		
		
		0V Calibration Value
		
		
		-199 -- +199  Internal Reference,Please Be Careful
		
		
		 
		
	
	
		
		F008
		
		
		+10V Calibration Value
		
		
		-199 -- +199  Internal Reference,Please Be Careful
		
		
		 
		
	
	
		
		F100
		
		
		Point Mode Num
		
		
		0: Point Mode Not Use   2-9: Point Num
		
		
		0
		
	
	
		
		F101…F109
		
		
		Point Output Value
		
		
		range : -10V to 10V  You can set as many values as there are points
		
		
		 
		
	




Current setting parameters. please read section 3.4 before operating, especially to enter a password:




	
		
		Number
		
		
		Description
		
		
		Note
		
		
		Default
		
	
	
		
		F001
		
		
		Adjust Mode
		
		
		0:Coarse  1:Fine  2: Point Mode(Need to set F100>0)
		
		
		0
		
	
	
		
		F002
		
		
		Output Mode
		
		
		0:0-20mA  1:4-20mA  2:0-22mA
		
		
		0
		
	
	
		
		F003
		
		
		Display Mode
		
		
		0:Real Voltage  1:Percentage 0-100.0%  2:50HZ
		
		
		0
		
	
	
		
		F004
		
		
		Add Or Sub Num For Knob’s Pulse(Coarse)
		
		
		1-50

		No Decimal Point Concept  (1-50)×10
		
		
		1
		
	
	
		
		F005
		
		
		Add Or Sub Num For Knob’s Pulse ( Fine)
		
		
		1-50

		No Decimal Point Concept  (1-50)×1
		
		
		1
		
	
	
		
		F008
		
		
		Calibration Value
		
		
		-999 -- +999  Internal Reference,Please Be Careful
		
		
		 
		
	
	
		
		F100
		
		
		Point Mode Num
		
		
		0: Point Mode Not Use   2-9: Point Num
		
		
		0
		
	
	
		
		F101…F109
		
		
		Point Output Value
		
		
		range : 0-22mA  You can set as many values as there are points
		
		
		 
		
	




current problem detection: the possibility of the current not output is very small,you can measured the voltage  of 4-AIO  6-GND with a multimeter, if it is >20V, than current file of your multimeter is broken or gear is wrong, because there are voltage must be current, try change a multimeter, multimeter current file inside a fuse is easy to burn;
Please Read This Manual Carefully Before Wiring,Turn Off The Power And Then Connect;